HEATING COOLING MACHINE AVAILABLE FOR HIRE
ATEX, FLP MODELS AVAILABLE

Select your Model

Please filter by temperature based on your specifications

SUNDI -25℃~200℃

Download PDF
ModelSUNDI-225WSUNDI-235WSUNDI-255WSUNDI-275WSUNDI-2A10WSUNDI-2A15WSUNDI-2A25WSUNDI-2A38WSUNDI-2A60WSUNDI-2A95WSUNDI-2A130WSUNDI-2A200W
Heating Power (kW)2.53.55.57.5101525386095130200
Cooling capacity at 200oC (kW)2.53.55.57.5101525386095130200
Cooling capacity at 100oC (kW)2.53.55.57.5101525386095130200
Cooling capacity at 50oC (kW)2.53.55.57.5101525386095130200
Cooling capacity at 20oC (kW)2.53.55.57.5101525386095130200
Cooling capacity at -5oC (kW)234.56.68121930467080125
Cooling capacity at -20oC (kW)11.82.83.84.67121622324365
Circulation Pump
Max Flow (l/min)
2035355075110150200250400400600
Circulation Pump
Max Pressure (bar)
22222.52.52.52.52.52.52.52.5
Connection sizeG1/2G3/4G3/4G1G1G1DN32DN40DN40DN50DN65DN65
Water-cooled type
W (AT 20℃)
600L/H800L/H1000L/H1200L/H1600L/H2200L/H6m³/H10m³/H14m³/H20m³/H28m³/H 36m³/H
Dimension cm50*85*130 55*100*17555*68*14555*100*17570*100*17570*100*175 80*120 *185100*95*185205*145*205205*145 *205245*145 *2250275*160 *225
Dimension(W) cm55*68 *14555*68 *14555*68*14555*100*17570*100*17570*100*175100*150 *185/////
Isolated ex-proof
size cm
45*110 *13045*110 *13055*120 *17055*120 *17070*125 *17580*145 *185105*120 *185100*95*185205*145*205 205*145 *205205*145 *205275*160 *225
Positive Pressure (W) cm110*95 *195110*95 *195110*95 *195110*95 *195110*95 *195135*145 *195(135+60) *145*225(135+60) *145*225(185+60) *145*225(185+100) *145*225(185+100) *145*225(185+150) *165*225
Weight kg1151651852302803005507501000 125015802150
Power MAX
380V 50HZ
AC 220V 4kW 6kW 8kW11kW14kW21kW34kW52kW79kW123kW 165kW 260kW
Optional size/50*68*14550*68*14550*68*145///////
  • Temp Range
  • -25oC~+200oC
  • Control Mode
  • Feedback PID + Our special dynamic control calculation, PLC controller
  • Temp Control
  • Process temp. control and jacket temp. control model
  • Temp Difference
  • Set or control the temperature difference between jacket oil and raw material process
  • Program Editor
  • 10 programs, each program can edit 40 steps.
  • Communication Protocol
  • MODBUS RTU Protocol,RS485 interface
  • Material Temp Feedback
  • PT100 OR 4~20mA or communication(normal: PT100)
  • Temp Feedback
  • The temp. of three points: the inlet and outlet of equipment,reactor material temperature(external temp. sensor)
  • Medium Temp Acccuracy
  • ±0.5oC
  • Material Temp Accuracy
  • ±1oC
  • Expansion Valve
  • Emerson thermal expansion valve/DANFOSS thermal expansion valve.
  • Evaporator
  • DANFOSS/KAORI Plate heat exchanger
  • Operation Panel
  • 7-inch touch screen, show temp. curve\EXCEL data output
  • Safety Protection
  • Self-diagnosis function; freezer overload protection; high pressure switch, overload relay, thermal protection device, liquid low level protection, high temperature protection and temp. fault protection.
  • Closed Circulation System
  • The whole system is full closed circulation, there is no oil mist at high temp. and no water vapor at low temp., pressure do not rise up when system is running. The system will supply oil automatically at low temperature.
  • Refrigerant
  • R-404A/R507C, optional: R448A

MANUFACTURED BY

CONTACT US

We automate and customize based on your specific requirement. Connect with us.

Phone +91 9618595672
Email info@e3techsys.com
Head Office 30-15-189, 4th Floor, Fonta Plaza,
Dabagardens, Visakhapatnam - 530020
Western Region No.602, 6th Floor, World Trade Centre
Sayajiganj, Vadodara-390005
Southern Region 111/314 C (X/269), Gokulam , Ayani Road
Maradu, Kochi - 682304